한국생산제조학회 학술지 영문 홈페이지
[ Papers ]
Journal of the Korean Society of Manufacturing Technology Engineers - Vol. 29, No. 2, pp.120-126
ISSN: 2508-5107 (Online)
Print publication date 15 Apr 2020
Received 19 Mar 2020 Revised 13 Apr 2020 Accepted 14 Apr 2020
DOI: https://doi.org/10.7735/ksmte.2020.29.2.120

교대 부양 욕창 방지 매트리스의 정량적 측정과 비교 방안

곽동기a ; 손동섭b ; 서갑호a, *
Objective Measurement and Comparison of Alternating Pressure Mattresses for Bed Sore Prevention
Dong-Gi Gwaka ; Dong-Seop Sohnb ; Kap-Ho Seoa, *
aInteraction Robotics R&D Division, Korea Institute of Robotics & Technology Convergence
bPlaning & Coordination Department, Korea Institute of Robotics & Technology Convergence

Correspondence to: *Tel.: +82-54-240-2501 E-mail address: neoworld@kiro.re.kr (Kap-Ho Seo).

Abstract

Alternating pressure mattresses are generally used for bed sore prevention as they inhibit the formation of pressure ulcers. However, several alternating pressure mattresses do not provide sufficient information, such as their numerical pressure, making it difficult for consumers to compare products. In this study, we propose a method for the objective measurement and comparison of alternating pressure mattresses. We measure the body pressure using a body pressure distribution sensor and obtain different frames depending on the air pressure state of the mattress. The frames are categorized into three types depending on the pressure ulcer factor. Although the patterns, shapes, and count of the mattresses are alternating, this method aids in comparing the performance of various mattresses. Furthermore, this measurement and data processing method can be applied to bed sore cushions.

Keywords:

Bed sore, Pressure ulcer, Alternating pressure mattress, Body pressure distribution sensor, Mattress comparison

Acknowledgments

본 연구는 산업통상자원부 산업기술혁신사업 ‘인공지능 기반 욕창 예방을 위한 자율형 침대 로봇 개발(20005004)’의 지원을 받아 수행된 연구임.

References

  • Pugliese, G., Favero, M. S., 1999, Pressure Ulcers Impact on Hospital Costs and Length of Stay, Infection Control & Hospital Epidemiology, 20:7 472. [https://doi.org/10.1017/S0195941700068843]
  • Department of Social Statistics Planning in Korea, 2017, viewed 12 Feburary 2020, Statistics on the Elderly in 2017, <http://kostat.go.kr/portal/korea/kor_nw/1/1/index.board?bmode=read&aSeq=363362, >.
  • Medical Resource Policy, 2019, viewed 16 Feburary 2020, Ministry of Health and Wlefare, Collection of Basic Data for Training High Quality Medical Personnel, <http://www.mohw.go.kr/react/al/sal0301vw.jsp?PAR_MENU_ID=04&MENU_ID=0403&page=1&CONT_SEQ =351956, >.
  • Lindgren, M., Unosson, M., Fredrikson, M., Ek, A. C., 2004, A Major Risk Factor for Development of Pressure Ulcers Among Adult Hospitalized Patients: A Prospective Study, Scandinavian Journal of Caring Science, 18:1 57-64. [https://doi.org/10.1046/j.0283-9318.2003.00250.x]
  • Kwon, J. W., Shim, C. S., Park, S. H., Lim, D. J., Lee, N. G., Shim, J. K., Hong, J. H., 2002, Kinematic Synthesis of a Medical Bed for Decubitus Ulcer Patients, The Korean Society of Mechanical Engineers, 1305-1310.
  • Lee, Y. D., Choi, M. S., Jang, I. H., Kim, C. Y., Choi, D. S., Kim, M. S., Kim, W. J., Kim, D. H., 2018, Research on the Development of Automated Multifunction-Integrated Motion Bed, The Journal of the Institute of Internet, Broadcasting and Communication, 18:5 215-222.
  • Tanimoto, Y., Takechi, H., Tokuhiro, A., Takechi, H., Yamamoto, H., 2002, Buttock Pressure Measurement using Bed Mattress, MTC/2002, Proceedings of the 19th IEEE Instrumentation and Measurement Technology Conference (IEEE Cat. No.00CH37276), 7803-7218. [https://doi.org/10.1109/IMTC.2002.1007112]
  • Lee, A. R., Jang, K. B., 2011, Development of The Physical Pressure Monitoring System to Prevent Pressure Ulcers, IEMEK Journal of Embedded Systems and Applications, 6:4 209-214.
  • Kim, C. J., Park, S. J., Kwon, K. S., Kim, J. S., Min, B. C., Lee, H. Y., Kim, K. H., 1999, The Characteristics of Body Pressure Distribution on Sleeping Posture, Science of Emotion & Sensibility, 2:1 95-103.
  • Jung, S. B., Hwang, I. H., Jung, J. S., Yuk, S. W., 2019, A Study on the Peak Pressure Measurement Test Method for Bed Robots for Preventing Bedsores, Proceedings of the KSMPE Autumn Conference, 274-274.
  • Lee, H., Cho, K. H., 2009, Changes on Blood Perfusion and Histological Structures of Sacral Skin in Rat by Externally Applied Pressure, Tissue Engineering and Regenerative Medicine, 6:1 294-299.
  • Cho, K. H., Yune, S. H., Lee, H., 2010, Effects of Mattresses and Position Positioning on Interface Pressure and Skin Blood Flow, Journal of Korean Academy of Rehabilitation Medicine, 34:2 214-219.