한국생산제조학회 학술지 영문 홈페이지
[ Article ]
Journal of the Korean Society of Manufacturing Technology Engineers - Vol. 33, No. 1, pp.50-57
ISSN: 2508-5107 (Online)
Print publication date 15 Feb 2024
Received 12 Jan 2024 Revised 02 Feb 2024 Accepted 06 Feb 2024
DOI: https://doi.org/10.7735/ksmte.2024.33.1.50

휴대용 안압 측정을 비전 센서 기반 정렬 시스템 개발

권순환a ; 김루나a ; 신소진a ; 이우영a ; 강진희b ; 장웅기a ; 김현욱a ; 임광석a ; 주창화a ; 김병희a, * ; 박희원c, *
Development of a Vision Sensor-based Alignment System for Portable Intraocular Pressure Measurement
Soonhwan Kwona ; Luna Kima ; SoJin Shina ; Uyeong Leea ; Jinhee Kangb ; Woongki Janga ; Hyunouk Kima ; Kwang Suk Lima ; Changhwa Jooa ; Byeonghee Kima, * ; Heewon Parkc, *
aDepartment of Smart Health Science and Technology, Kangwon National University
bC&VTech Co., Ltd.
cDepartment of Rehabilitation Medicine, Kangwon National University Hospital

Correspondence to: *Tel.: +82-33-250-6374 E-mail address: kbh@kangwon.ac.kr (Byeonghee Kim). Correspondence to: *Tel.: +82-33-248-7701 E-mail address: HWPARK@kangwon.ac.kr (Heewon Park).

Abstract

Elevated intraocular pressure(IOP) is a primary cause of glaucoma, rendering IOP measurement essential for diagnosis. Traditional IOP measurement devices are bulky and stationary, necessitating movement between consultation and measurement rooms. Portable tonometers address this issue but face challenges in ensuring proper alignment between the cornea and the device for precise measurements. This study proposes a method enabling medical professionals or third parties to determine this alignment using a portable tonometer. The proposed alignment system is designed considering structural compactness and cost competitiveness and comprises three light sources and vision sensors. To validate the proposed system, experiments were conducted, proving its ability to determine the alignment between the cornea and the system.

Keywords:

Active senior, Glaucoma, Intraocular pressure, Non-contact tonometer, Alignment system

Acknowledgments

본 연구는 2024년도 교육부의 재원으로 한국연구재단의 지원을받아 수행된 지자체-대학 협력기반 지역혁신 사업의 결과입니다(2022RIS-005).

References

  • Kim, H. W., Zhang, X., Kim, Y. S., Jung, I. H., 2022, Comparison of the Performance of CNN Models for Retinal Diseases Diagnosis, Journal of Korean Institute of Intelligent Systems, 32:1 51-60. [https://doi.org/10.5391/JKIIS.2022.32.1.51]
  • Lee, S. K., Kim, S. K., 2005, Detection of the Optic Disk Boundary in Retinal Images Using Inward and Outward Curve Evolution, The Journal of the Korea Contents Association, 5:6 138-145.
  • Kim, M. S., Woo, S. Y., Lee, S., Park, Y. K., 2007, Tonometry and Its Standards, J. Korean Soc. Precis. Eng., 24:9 17-22.
  • Tojo, N., Oka, M., Miyakoshi, A., Ozaki, H., Hayashi, A., 2014, Comparison of Fluctuations of Intraocular Pressure before and after Selective Laser Trabeculoplasty in Normal-tension Glaucoma Patients, J. Glaucoma, 23:8 138-143. [https://doi.org/10.1097/IJG.0000000000000026]
  • Collins, R., Van der Werff, T. J.., 2013, Mathematical Models of the Dynamics of the Human Eye, Springer, New York.
  • Goel, M., Picciani, P. G., Lee, R. K., Bhattacharya, S. K., 2010, Aqueous Humor Dynamics: A Review, Open Ophthalmol. J., 4 52-59. [https://doi.org/10.2174/1874364101004010052]
  • Stevens, S., Gilbert, C., Astbury, N., 2007, How to Measure Intraocular Pressure: Applanation Tonometry, Community Eye Health, 20:64 74-75.
  • Northrop, R. B., 2002, Noninvasive Instrumentation and Measurement in Medical Diagnosis, CRC Press, Boca Raton.
  • Touboul, D., Roberts, C., Kerautret, J., Garra, C., Maurice-Tison, S., Saubusse, E., Colin, J., 2008, Correlations between Corneal Hysteresis, Intraocular Pressure, and Corneal Central Pachymetry, J. Cataract. Refract. Surg., 34:4 616-622. [https://doi.org/10.1016/j.jcrs.2007.11.051]
  • Nakao, Y., Kiuchi, Y., Okimoto, S., 2017, A Comparison of the Corrected Intraocular Pressure Obtained by the Corvis ST and Reichert 7CR Tonometers in Glaucoma Patients, PLoS One, 12:1 e0170206. [https://doi.org/10.1371/journal.pone.0170206]
  • Ulvik, S. O., Seland, J. H., Wentzel-Larsen, T., 2005, Refraction, Axial Length and Age-related Maculopathy, Acta Ophthalmol. Scand., 83:5 419-423. [https://doi.org/10.1111/j.1395-3907.2005.520_corr.x]
  • Fledelius, H. C., 1988, Refraction and Eye Size in the Elderly: A Review based on Literature, Including Own Results, Acta Ophthalmol., 66:3 241-248. [https://doi.org/10.1111/j.1755-3768.1988.tb04592.x]
  • Piñero, D. P., Puche, A. B. P., Alió, J. L., 2008, Corneal Diameter Measurements by Corneal Topography and Angle-to-angle Measurements by Optical Coherence Tomography: Evaluation of Equivalence, J. Cataract. Refract. Surg., 34:1 126-131. [https://doi.org/10.1016/j.jcrs.2007.10.010]
  • Khng, C., Osher, R. H., 2008, Evaluation of the Relationship between Corneal Diameter and Lens Diameter, J. Cataract. Refract. Surg., 34:3 475-479. [https://doi.org/10.1016/j.jcrs.2007.10.043]
  • Kim, H. J., 2016, Development of Non-contact type Intraocular Pressure Measurement System based on Rebounding Sound and Air-pressure, Doctorate Thesis, Kangwon National University, South Korea.
Soonhwan Kwon

Ph.D. Candidate in the Department of Smart Health Science and Technology, Kangwon National University. His research interest is Precision Control of Feed Derive System.

E-mail: didantm@kangwon.ac.kr

Luna Kim

Master Candidate in the Department of Smart Health Science and Technology, Kangwon National University. Her research interest is Cancer Immunotherapy.

E-mail: luna1223@kangwon.ac.kr

SoJin Shin

Master Candidate in the Department of Smart Health Science and Technology, Kangwon National University. Her research interest is Nano Bio Engineering.

E-mail: tlsthwls47@kangwon.ac.kr

Uyeong Lee

Master Candidate in the Department of Smart Health Science and Technology, Kangwon National University. His research interest is Physical Training.

E-mail: koreahoke@naver.com

Jinhee Kang

President in C&VTech Co., Ltd.. His research interest is Eye Health.

E-mail: ceo@cnvtech.co.kr

Woongki Jang

Post-doctoral Associate in the Department of Smart Health Science and Technology, Kangwon National University. His research interest is Micro/Nano Scale Surface Texturing Technologies and Digital Twin.

E-mail: wkddndrl@kangwon.ac.kr

Hyunouk Kim

Professor in the Department of Smart Health Science and Technology, Kangwon National University. His research interest is Nano Bio Engineering.

E-mail: kimhoman@kangwon.ac.kr

Kwang Suk Lim

Professor in the Department of Smart Health Science and Technology, Kangwon National University. His research interest is Drug Delivery System.

E-mail: kslim@kangwon.ac.kr

Changhwa Joo

Professor in the Department of Smart Health Science and Technology, Kangwon National University. His research interest is Physiology & Physical Training.

E-mail: phd.football@gmail.com

Byeonghee Kim

Professor in the Department of Smart Health Science and Technology, Kangwon National University. His research interest is Micro and Nano System Design and Precision Control of Machine Tools.

E-mail: kbh@kangwon.ac.kr

Heewon Park

Physician in the Department of Rehabilitation Medicine, Kangwon National University Hospital. His research interest is Rehabilitation Medicine.

E-mail: HWPARK@kangwon.ac.kr